Biosafety Cabinets (BSCs): Types, Selection, and Best Practices 🦺

Biosafety Cabinets (BSCs) are critical laboratory devices designed to provide a controlled environment for handling infectious agents and hazardous materials. These cabinets are integral to maintaining laboratory safety by preventing the escape of harmful agents, thus protecting both the user and the environment. BSCs are widely used in research, clinical, and pharmaceutical laboratories to ensure that biohazardous materials are contained and managed safely.

Types of Biosafety Cabinets

Class I BSCs

Class I BSCs provide personnel and environmental protection but do not offer product protection. They have an inward airflow to protect the user and HEPA-filtered exhaust air to protect the environment. These cabinets are suitable for work with low to moderate-risk agents where product sterility is not a concern.

Class II BSCs

Class II BSCs are the most commonly used cabinets in laboratories. They offer both product and environmental protection and are divided into four types:

  • Type A1: Recirculates 70% of the air within the cabinet through HEPA filters, suitable for work with non-volatile toxic chemicals and radionuclides.
  • Type A2: Similar to Type A1 but with higher face velocity, providing increased protection.
  • Type B1: Exhausts 70% of the air through HEPA filters, used for work with volatile toxic chemicals in small amounts.
  • Type B2: 100% of the air is exhausted through HEPA filters, ideal for work with volatile toxic chemicals and radionuclides.

Class II BSCs are essential in clinical, pharmaceutical, and research laboratories due to their comprehensive protection capabilities.

Class III BSCs

Class III BSCs, also known as glove boxes, provide the highest level of protection. They are completely enclosed, and operations within the cabinet are conducted through attached gloves. These cabinets are used for high-risk pathogens and bioweapons research, ensuring maximum containment.

Applicable International Standards

  • ISO 14644: Cleanrooms and associated environments.
  • ISO 13485: Quality management systems for the design and manufacture of medical devices.
  • NSF/ANSI 49: Specific to biosafety cabinetry performance in the U.S.
  • IEC 61010-1: Safety requirements for electrical equipment for measurement, control, and laboratory use.

Design Considerations

Biosafety Cabinets (BSCs) are essential for ensuring the safety of personnel, products, and the environment in laboratories handling biological materials. This guide delves deeper into the design considerations, manufacturing requirements, and validation processes to meet the highest safety and quality standards.


BSCs are categorized into three primary classes based on their protection capabilities and intended applications.

Class I:

  • Primary Function: Protect personnel and the environment from exposure to hazardous agents.
  • Limitation: Does not provide product protection, making it unsuitable for sterile procedures.
  • Application: Handling low-risk biological materials, such as teaching laboratories or procedures involving non-sterile materials.

Class II:

  • Primary Function: Provides comprehensive protection for personnel, product, and the environment.
  • Subtypes:
    • Type A1/A2:
      • Recirculates a significant portion of filtered air back into the cabinet, with partial exhaust to the room or an external exhaust system.
      • Suitable for general microbiological work without volatile chemicals.
    • Type B1/B2:
      • Type B1: Recirculates a minimal portion of air, with most air exhausted through HEPA filters.
      • Type B2: 100% exhausts filtered air, making it ideal for procedures involving volatile chemicals or radionuclides.

Class III:

  • Designed For: Maximum containment of highly hazardous biological agents (e.g., biosafety level 4 laboratories).
  • Key Features: Airtight design with a glovebox structure, ensuring no direct contact with the workspace.

Key Design Features

Airflow Dynamics:

  • Laminar (Unidirectional) Airflow: Ensures controlled and uniform airflow, minimizing turbulence that could disrupt containment or contaminate products.
  • Containment through HEPA/ULPA Filters:
    • HEPA filters trap ≥99.99% of particles ≥0.3 microns.
    • Ultra-Low Penetration Air (ULPA) filters may be used for even higher efficiency, meeting ISO and EN standards.

HEPA/ULPA Filters:

  • Filtration Efficiency: Filters must comply with EN 1822 standards or equivalents like NSF/ANSI 49.
  • Multi-Stage Filtration: Secondary filters provide redundancy, particularly in Class III cabinets.

Cabinet Structure:

  • Internal Components: Constructed with stainless steel (SS 304/316), offering durability, corrosion resistance, and ease of decontamination.
  • External Housing: Powder-coated steel or composite materials are used for robustness and resistance to environmental factors.
  • Ergonomics: Features such as sloped sashes, adjustable armrests, and spacious work areas reduce operator fatigue and enhance usability.

Containment and Seals:

  • Airtight Construction: Essential for Class III BSCs, ensuring no air leakage under operational or idle conditions.
  • Gasketed Seals: Used on access doors and panels to maintain containment integrity.

Ventilation Systems:

  • Integrated Monitoring: Alarms alert users to disruptions in airflow or filter saturation.
  • UV Sterilization: Optional systems, compliant with ISO 15858, provide an additional layer of sterilization when the cabinet is idle.

Energy Efficiency:

  • Use of low-energy fans, LED lighting, and optimized motor systems ensures reduced operational costs and environmental impact.

Noise Control:

  • Design Goals: Cabinets are engineered to operate below 65 dB, minimizing disruption to operators while maintaining compliance with occupational safety standards.

Safety and Ergonomics

Fail-Safe Mechanisms:

  • Automatic shutdown of non-essential systems during power loss to prevent containment breaches.

Interlocked Sashes:

  • Prevent sashes from being opened in a manner that compromises containment, ensuring operational safety.

Lighting:

  • Adequate illumination (≥1000 lux) is critical for precision and reduces strain on operators during detailed procedures.

Manufacturing Requirements

The manufacturing of Biosafety Cabinets (BSCs) requires attention to material selection, precision engineering, and quality control processes to ensure safety, durability, and regulatory compliance;

Material Selection

The choice of materials directly impacts the BSC's performance, longevity, and compatibility with operational requirements.

Internal Chambers

  • Material: Stainless steel (SS 304 or SS 316).
    • Properties of SS 304:
      • High resistance to corrosion and rust under typical laboratory conditions.
      • Economical option suitable for most biosafety applications.
    • Properties of SS 316:
      • Enhanced resistance to pitting and chemical attack, especially in high-humidity environments or when exposed to harsh decontaminants (e.g., chlorine-based agents).
      • Often used in high-containment cabinets (Class III) or specialized applications.
  • Rationale:
    • Smooth, non-porous surfaces prevent microbial adhesion and ensure compatibility with cleaning protocols.
    • Withstands repeated exposure to disinfectants and sterilizing agents without degradation.

External Housing

  • Material: Powder-coated steel or composite materials.
    • Powder Coating Features:
      • Provides a durable finish resistant to scratches, dents, and environmental factors such as humidity or chemical exposure.
      • Available in anti-microbial formulations to further enhance safety.
  • Composite Materials:
    • Lightweight alternatives (e.g., polymer composites) may be used for portability or specialized applications.
    • Fire-retardant properties enhance safety in laboratory settings.

Transparent Panels

  • Material: Laminated safety glass with UV resistance.
    • Key Features:
      • Protects operators from UV exposure during sterilization cycles in cabinets with UV systems.
      • Maintains structural integrity under high-stress conditions, such as sudden impacts or pressure changes.
      • Ensures long-term optical clarity for visibility into the workspace.

Manufacturing Processes

The manufacturing processes for BSCs involve high-precision engineering to meet stringent safety and containment standards.

Precision Welding

  • Purpose: Ensures the cabinet's structural integrity and airtight containment.
  • Process:
    • Automated or semi-automated TIG (Tungsten Inert Gas) welding is commonly used for stainless steel components.
    • Welds are inspected using non-destructive testing methods, such as dye penetrant or ultrasonic testing, to detect cracks or voids.
  • Benefits:
    • Eliminates potential leaks in seams, which is critical for Class III cabinets handling high-risk pathogens.
    • Contributes to laminar airflow consistency by maintaining smooth interior surfaces.

Surface Treatments

  • Objective: Enhance the durability and cleanliness of internal and external surfaces.
  • Processes:
    • Polishing:
      • Internal surfaces are polished to a mirror-like finish to minimize microbial adherence.
      • Mechanical or electro-polishing techniques may be used, with electro-polishing providing superior smoothness at the microscopic level.
    • Coatings:
      • Anti-microbial coatings may be applied to external surfaces to inhibit microbial growth.
      • Corrosion-resistant coatings ensure long-term durability under harsh conditions.
  • Rationale:
    • Treated surfaces facilitate easy decontamination and reduce the risk of cross-contamination.

Filter Integration

  • Purpose: Proper integration of HEPA and ULPA filters ensures effective containment and air purification.
  • Process:
    • Filters are installed in sealed housings designed to eliminate bypass airflow.
    • Compliance with ISO 14644-1 cleanroom standards ensures the filter environment is free from particulates during installation.
    • Gaskets and sealants are applied to prevent leaks and maintain pressure integrity.
  • Testing:
    • Each filter assembly undergoes in-situ integrity testing (e.g., using PAO or DOP aerosols) to confirm containment before final cabinet assembly.
    • Continuous pressure monitoring systems are integrated to alert users of filter saturation or damage.

Additional Manufacturing Considerations

Ergonomics in Assembly

  • Components such as sloped sashes, armrests, and work surface heights are incorporated during the assembly stage to ensure operator comfort.

Automation and Robotics

  • Advanced manufacturers use robotic systems for cutting, welding, and assembling components to improve precision and repeatability.

Quality Assurance

  • Every stage of manufacturing is subject to stringent quality control measures, including dimensional inspections, material certification, and process audits.

Testing and Validation

Testing and validation are critical to ensuring that Biosafety Cabinets (BSCs) meet stringent performance, safety, and ergonomic standards. These tests verify containment efficiency, operational reliability, and compliance with international regulations. 

Factory Acceptance Tests (FAT)

Factory Acceptance Testing (FAT) is a quality assurance process conducted before shipping the BSCs to end users. It ensures that the cabinet meets design specifications and operational criteria.

1. Airflow Tests:
Airflow uniformity is vital for maintaining containment and minimizing contamination risks, to confirm that the cabinet provides consistent laminar airflow across the workspace without turbulence.

  • Procedure:
    • Use an anemometer or thermal airflow sensor to measure airflow velocity at multiple points within the cabinet.
    • Compare measured values against specified ranges (e.g., 0.3–0.5 m/s for Class II BSCs).
    • Ensure airflow uniformity, with deviations not exceeding specified limits (e.g., ±20% deviation across measurement points).
  • Acceptance Criteria: Airflow patterns should be consistent, with no reverse or stagnant zones that could compromise containment.

2. Filter Integrity Tests:
HEPA or ULPA filters are critical for trapping particulates and pathogens. Any compromise in filter integrity can lead to containment failure.

  • Procedure:
    • Introduce a challenge aerosol (e.g., poly-alpha-olefin [PAO] or dioctyl phthalate [DOP]) upstream of the filter.
    • Use a photometer or particle counter downstream to detect aerosol penetration.
    • Scan the filter surface and housing seals for leaks.
  • Acceptance Criteria: Penetration levels should not exceed 0.01% for HEPA filters, ensuring compliance with ISO 14644-3 and EN 12469 standards.

3. Containment Tests:
Containment integrity is paramount, especially for BSCs handling hazardous materials. It is done to verify that the cabinet’s structure prevents air leakage under operational conditions.

  • Procedure:
    • Pressurize the cabinet’s internal chamber and monitor for pressure decay using calibrated equipment.
    • Conduct smoke or tracer gas testing to visualize airflow patterns and detect potential leaks.
  • Acceptance Criteria: No detectable leaks should be observed during testing. Pressure decay values must remain within allowable limits.

Microbial Challenge Testing:

Microbial challenge testing evaluates the cabinet’s ability to contain biological hazards under simulated real-world conditions. It is  done to validate the containment and filtering capability of the BSC when exposed to live microorganisms.

  • Procedure:
    • Introduce a biological indicator, such as Bacillus spores, into the cabinet’s airflow system.
    • Measure the containment efficiency of HEPA/ULPA filters by detecting the presence of spores downstream.
    • Repeat the process for different operational scenarios, including simulated breaches.
  • Acceptance Criteria: No spores should bypass the filtration system, demonstrating the cabinet’s effectiveness in containing biohazards.

Noise, Vibration, and Lighting Tests:

Ensuring a safe and comfortable working environment is as important as containment. Noise, vibration, and lighting levels directly impact operator efficiency and safety.

1. Noise Tests: To ensure the BSC operates within acceptable noise levels for user comfort.

  • Procedure:
    • Measure sound levels using a decibel meter positioned at the operator’s ear level.
    • Conduct tests under standard operating conditions with the cabinet fully functional.
  • Acceptance Criteria: Noise levels should not exceed 65 dB, in line with ISO 11201 ergonomic standards.

2. Vibration Tests: To confirm that cabinet vibrations do not interfere with delicate procedures or compromise stability.

  • Procedure:
    • Place sensitive instruments (e.g., accelerometers) on the work surface to measure vibration amplitude and frequency.
    • Test under different airflow and operational settings.
  • Acceptance Criteria: Vibration levels must remain minimal, ensuring no disturbance to precision equipment or operator performance.

3. Lighting Tests: To ensure sufficient illumination for laboratory tasks without glare or shadows.

  • Procedure:
    • Measure light intensity using a lux meter at various points within the workspace.
    • Check for uniform distribution of light across the work surface.
  • Acceptance Criteria: Illumination levels should meet or exceed 1000 lux, ensuring clear visibility for precision work.

Regulatory Considerations and Choice of Biosafety Cabinets for Various Applications

Selecting the appropriate Biosafety Cabinet (BSC) is a critical decision influenced by regulatory standards, safety requirements, and the specific applications for which the cabinet will be used. Regulatory guidelines, such as those from NSF/ANSI Standard 49 and EN 12469, provide the framework for ensuring that BSCs meet performance, containment, and airflow criteria necessary for protecting both personnel and products.

Another important consideration is the work zone environment. Cleanroom classifications (ISO 3 to ISO 9) dictate the level of cleanliness and airflow control required, directly impacting the choice of cabinet. For example, pharmaceutical manufacturing or sensitive semiconductor work may require BSCs that conform to ISO Class 5 standards.

Lastly, facilities handling hazardous or volatile substances must choose BSCs designed for chemical containment, such as Class II Type B2 or Class III cabinets, which ensure proper ducting and airflow to prevent exposure.

Classification of Biological Safety Cabinets (per NSF / ANSI Standard 49)

Class

 Inflow Velocity (m/s)

 Recirculating Air (%)

 Exhaust Air (%)

 Metal Plenum Surrounded by

 Exhaust Alternatives

Biosafety Level

I **

0.38

0

100

Outside Air

Inside Room/ Hard Duct

1,2 & 3

II Type A1

0.38

70

30

Outside Air

Inside Room/ Thimble Duct

1,2 & 3 *

II Type A2 **

0.50

70

30

Negative Plenum

Inside Room/ Thimble Duct

1,2 & 3 *

II Type B1

0.50

30

70

Negative Plenum

Hard Duct only

1,2 & 3 *

II Type B2

0.50

0

100

Negative Plenum

Hard Duct only

1,2 & 3 *

III **

Closed: > 0.5” WC

0

100

Negative Plenum

Inside Room/ Hard Duct

1,2, 3 & 4

* Open front cabinets (e.g. Class II BSC) can still be used in BSL 4 facilities but will require positive-pressured personnel suit for laboratory users. 

** EN 12469 only recognizes Class I, Class II and Class III BSCs; Class II BSC Inflow velocity requirement as per EN 12469: ≥0.40 m/s

Applications with work zone ISO Class requirement

ISO Class

Application

3 / 4

Sensitive semiconductor or pharmaceutical work

5

Laminar flow cabinets, biological safety cabinets, pharmaceutical isolators

7

Clean room for pharmaceutical preparation, used with laminar flow or biosafety cabinets inside

8

Typical hospital environment

9

Typical office or labs without air filtering

Selection of a Safety Cabinet through Risk Assessment

A BSC should be selected primarily according to the type of protection required: product protection; personnel protection against Risk Group 1–4 microorganisms; personnel protection against exposure to radionuclides and volatile toxic chemicals; or a combination of these. Below table shows which BSCs are recommended depending on the necessary type of protection;

Type of Protection

BSC Selection

Personnel Protection against RG 1-3 microorganisms

Class I, Class II, Class III BSC

Personnel protection against RG 4 microorganisms, glove-box laboratory

Class III BSC

Personnel protection against RG 4 microorganisms, suit laboratory

Class I, Class II BSC

Product Protection

Class II, Class III BSC only if laminar flow included

Volatile radionuclide/chemical protection (small amount)

Class II Type B1, Class II Type A2 (with thimble ducting) BSC

Volatile radionuclide/chemical protection

Class I, Class II Type B2, Class III BSC


Performance Testing BSCs in the Field

All Biological Safety Cabinets are to be verified to the current NSF/ANSI Standard 49, Annex F or EN12469 upon installation and annually thereafter. The purpose and acceptance level of the operational tests ensure the equalization of inflow and exhaust air, the circulation of air onto the work surface, and the integrity of the cabinet and the filters. Other tests monitor the BSCs electrical and physical features.

Downflow Velocity Test: This test determines the velocity of air traveling through the cabinet workspace.6 The EN12469 indicates a permissible downflow velocity range of 0.25-0.50m/s, whereas the NSF49 does not specify any downflow velocity requirement. A Thermo-anemometer shall be used to carry out the test in accordance with NSF49 while the EN12468 does not specify the test instrument accuracy and type.

Inflow Velocity Test: This test determines the average speed of air entering the cabinet. To carry out the test, direct inflow measurement (DIM) instrument shall be used to measure the inlet volumetric flow rate on the front aperture at nominal operating speed (primary method) and Thermal anemometers or pitot tubes or both shall be used to verify the calculated inflow velocity (secondary method) in accordance to NSF49. The European method for measuring the rate of inflow is performed above the exhaust filter. Both EN12469 and NSF49 specify minimum requirements for inflow velocity. For Class II Type A2 cabinets, the NSF49 specifies a minimum inflow rate of 0.5 m/s, whereas the European Standard requires 0.4 m/s for Class II cabinets.

Airflow Smoke Pattern Test: This test determines 

  • whether the movement of the air along the entire perimeter of the work access opening is inward 
  • whether the movement of the air within the working area is downward with no dead spots or refluxing 
  • whether ambient air flows onto or over the work surface 
  • whether airflow within the cabinet does not escape outside at the sides and top of the sash. A suitable smoke generator shall be used to visualize the cabinet’s airflow pattern.

HEPA/ULPA Filter Leak Test: This test determines the integrity of supply and exhaust HEPA filters, filter housing and filter mounting frames. To carry out the test, an aerosol generator shall be used to evenly distribute the aerosol throughout the supply (positive) cabinet plenum and an aerosol photometer shall be used to monitor aerosol penetration downstream of the filter, and to scan for the presence of leaks in accordance to NSF49. The European Standard allows an alternative test using the natural challenge test.

Light Intensity Test: This test determines the intensity of light on the work surface of the cabinet. A light intensity meter is used to obtain light levels at the cabinet work surface. NSF49 allows a slightly lower lighting level of 650 lux whereas the EN standard requires 750 lux.

Noise Level Test: This test determines the noise levels produced by the cabinets. With the cabinet running under regular parameters, a calibrated noise level meter shall be used to obtain a noise level of the cabinet during normal operation. The EN12469 test method specifies a distance further away from the cabinet as compared to the NSF49. The EN12469 allows up to 65dBa whereas the NSF49 allows up to of 67dBa.

Vibration Test: This test determines the amount of vibration in an operating cabinet. Vibration testing meter shall be used to verify the vibration level at the work tray

UV Radiation Intensity Test: This test determines the energy output of the UV lamp’s sufficiency in killing the microorganisms within the cabinet’s work zone. 70% ethanol shall be used to clean the surface of the bulb prior to performing the test. To carry out the test, UV radiation intensity meter shall be used to obtain light intensity at work surface level within the cabinet. UV Radiation intensity inside the cabinet should not be less than 40 μW/cm2 at 254 nanometers (nm).

Containment Test using KI Discus Method: The KI (potassium iodide) discus test is defined in the European Standard for microbiological safety cabinets, EN12469, as a test method for validating the operator protection capabilities of the cabinet. The KI Discus test has been designed to enable operator protection factors to be measured for class I and Class II open fronted biological safety cabinets. Unlike test methods employing a microbiological aerosol challenge this technique enables cabinets to be evaluated without the risk of microbial contamination of either the biological safety cabinet or the laboratory.

The selection and use of Biological Safety Cabinets (BSCs) are not merely technical choices but essential components of ensuring safety, compliance, and operational efficiency in laboratory environments. By understanding the classifications, regulatory standards, and specific application needs, organizations can effectively mitigate risks to personnel, products, and the environment. Whether for research, clinical diagnostics, or industrial applications, the right BSC not only safeguards work but also underscores a commitment to best practices in biosafety and quality assurance.