Material Characterization and Analytical Testing of Biomaterials

Biomaterials play a critical role in the development of medical devices and implants, impacting their safety, efficacy, and longevity. The characterization and analytical testing of biomaterials are essential steps in ensuring their suitability for biomedical applications.

Analytical procedures provide the initial means for investigating the biocompatibility of medical device materials. Knowledge of device materials and their propensity for releasing leachable matter will help manufacturers assess the risks of in vivo reactivity and preclude subsequent toxicology problems with finished devices. Increasingly, 
Regulatory Authorities has been asking for analytical characterization of device materials and potential leachables per ISO 10993-17 and 10993-18. Many manufacturers also use analytical procedures for routine QC of raw materials or finished products. The degree of chemical characterization required should reflect the nature and duration of the clinical exposure and should be determined based on the data necessary to evaluate the biological safety of the device. It will also depend on the nature of the materials used, e.g. liquids, gels, polymers, metals, ceramics, composites or biologically sourced material.


The following methodology is practiced in industry for chemical characterization of a device material:

  1. Determine the qualitative composition of each device component or material. This information should be available from the material vendor, or it can be determined through laboratory testing. The list of constituents should include
    1. the identity of the matrix (i.e. the major component such as the specific polymer, alloy, or metal) 
    2. all plasticizers, colorants, anti-oxidants, fillers, etc. deliberately added during fabrication of the material 
    3. impurities such as unreacted monomers and oligomers 
    4. manufacturing materials such as solvent residues, slip agents, and lubricants.
  2. Estimate the potential for patient exposure for each item on the material constituent list. Use literature searches of toxicological databases to assess the likelihood of tissue reactivity. For potentially toxic constituents, design and conduct laboratory studies to determine the extractable levels of those constituents. Use exaggerated conditions of time and temperature, and consider appropriate detection limits. Additional studies may be needed to assess levels of extractables released in actual use conditions.
  3. Data generated from this characterization process can be used to create a material data file. The information can then be used as a reference for continued testing of device materials to ensure consistency of future production lots. This may in turn reduce the need for routine biological testing.

Additional uses of analytical characterization data might include:

  1. Use in an assessment of the overall biological safety of a medical device. 
  2. Measurement of the level of any leachable substance in a medical device in order to allow the assessment of compliance with the allowable limit derived for that substance from health based risk assessment. 
  3. Judging equivalence of a proposed material to a clinically established material. 
  4. Judging equivalence of a final device to a prototype device to check the relevance of data on the latter to be used to support the assessment of the former. 
  5. Screening of potential new materials for suitability in a medical device for a proposed clinical application.

Traditional Extractable Material Characterization

Traditional extractable material characterization is a vital process in medical device manufacturing, focusing on identifying and analyzing substances that can be extracted from device materials under specific conditions. By subjecting samples to controlled extraction conditions and analyzing the resulting extracts using various analytical techniques, manufacturers can assess the safety and suitability of materials for use in medical devices. This characterization process informs decisions regarding material selection, regulatory compliance, and overall patient safety.

  • USP Physicochemical Tests for Plastics & USP Physicochemical Test Panel for Elastomeric Closures for Injections: These tests are designed to evaluate the physicochemical properties of elastomeric closures used in injections, ensuring their safety and compatibility with pharmaceutical products.
  • USP Polyethylene Containers Tests – Heavy Metals and Non-volatile Residues: Polyethylene containers undergo specific tests outlined by the US Pharmacopeia (USP) to assess the presence of heavy metals and non-volatile residues, ensuring compliance with regulatory standards for pharmaceutical packaging.
  • Indirect Food Additives and Polymers Extractables (21CFR Part 177): This regulation addresses the safety of indirect food additives, including polymers used in food packaging and processing equipment. It outlines requirements for assessing extractables from polymers to prevent contamination of food products.
  • Sterilant Residues – Ethylene Oxide, Ethylene Chlorohydrin, Ethylene Glycol: Residual sterilants such as ethylene oxide, ethylene chlorohydrin, and ethylene glycol can pose health risks if present in medical devices or pharmaceutical packaging. Testing for sterilant residues ensures that products meet safety standards and do not pose harm to patients or consumers.


Test procedures for extractable materials

Test procedures for extractable materials involve several key steps. First, samples of the material under consideration are prepared following specified guidelines to ensure representative sampling. These samples are then subjected to extraction using suitable solvents or conditions that mimic real-world exposure scenarios. Once extracted, the samples undergo analysis using various techniques such as chromatography and spectroscopy to identify and quantify any extractable substances present. The results are interpreted in light of regulatory requirements and industry standards to assess compliance and potential risks. Finally, a comprehensive report is generated, documenting the testing procedures, results, and conclusions, which serve as a basis for decision-making regarding the material's suitability for its intended application.

Test procedures for extractable materials often involve a variety of analytical techniques to detect and quantify substances present. These techniques include:
  1. UV/Visible Spectroscopy: Measures absorption of UV or visible light by molecules in the sample, providing information about the presence of specific compounds.
  2. Gas Chromatography (GC): Separates and analyzes volatile compounds in the sample, useful for identifying organic compounds.
  3. Liquid Chromatography (LC): Separates and analyzes non-volatile or less volatile compounds in the sample, complementing GC analysis.
  4. Infrared Spectroscopy (IR): Determines the presence of functional groups in molecules by measuring absorption of infrared radiation.
  5. Mass Spectrometry: Identifies compounds based on their mass-to-charge ratio, providing detailed information about molecular structure.
  6. Residual Solvents Analysis: Detects and quantifies solvents remaining in the material after manufacturing processes.
  7. Atomic Absorption Spectroscopy (AAS): Measures the concentration of specific elements in the sample based on the absorption of light at characteristic wavelengths.
  8. Inductively Coupled Plasma Spectroscopy (ICP): Determines the elemental composition of the sample by analyzing ionized atoms using plasma.
These techniques, used individually or in combination, enable thorough characterization of extractable materials, ensuring compliance with regulatory standards and assessing potential risks associated with their use in medical devices.


Bulk Material Characterization

Bulk material characterization involves the comprehensive analysis of the physical, chemical, and mechanical properties of raw materials used in the manufacturing of medical devices. This process aims to ensure the quality, consistency, and suitability of the materials for their intended applications. Key aspects of bulk material characterization include:
  • Chemical Composition: Determining the elemental composition and identifying any impurities or contaminants present in the material.
  • Physical Properties: Assessing characteristics such as density, porosity, particle size distribution, surface area, and morphology.
  • Mechanical Properties: Evaluating mechanical behavior under different conditions, including tensile strength, modulus of elasticity, hardness, and flexibility.
  • Thermal Properties: Analyzing thermal behavior, such as melting point, glass transition temperature, thermal conductivity, and coefficient of thermal expansion.
  • Electrical Properties: Measuring electrical conductivity, resistivity, dielectric strength, and other relevant electrical properties.
  • Biological Compatibility: Conducting biocompatibility tests to evaluate the material's interaction with living tissues and ensuring it meets regulatory requirements for safety.
By thoroughly characterizing bulk materials, manufacturers can make informed decisions during the design and production of medical devices, ensuring they meet quality standards, performance specifications, and regulatory requirements.

Bulk material characterization encompasses various analytical techniques to assess the properties of raw materials used in medical device manufacturing. These techniques include:
  • Infrared Spectroscopy Analysis: Utilized to determine the identity and estimate the gross composition of materials. This method includes both reflectance and transmission spectroscopy, offering insights into molecular structure and functional groups present in the material.
  • Atomic Absorption Spectroscopy (AAS): A technique used to quantify the concentration of specific elements present in the material. AAS is particularly useful for detecting trace metals and heavy metals.
  • Inductively-Coupled Plasma Spectroscopy (ICP): Another method for elemental analysis, ICP spectroscopy is highly sensitive and can detect a wide range of elements with excellent precision.
  • Thermal Analysis: This approach involves studying the thermal properties of materials, including melting point, glass transition temperature, and thermal stability. Techniques such as differential scanning calorimetry (DSC) and thermogravimetric analysis (TGA) are commonly employed for thermal characterization.
These analytical techniques provide valuable insights into the chemical composition, elemental content, and thermal behavior of bulk materials, aiding in quality control, material selection, and ensuring compliance with regulatory standards in medical device manufacturing.



Surface Characterization 

Surface characterization involves analyzing the outermost layer of materials used in medical device manufacturing. Key techniques for surface characterization include:
  • IR Reflectance Spectroscopy: Infrared (IR) reflectance spectroscopy is used to study the chemical composition and structure of the surface layer of materials. By measuring the interaction of infrared light with the material's surface, valuable information about surface functional groups and molecular bonds can be obtained.
  • Scanning Electron Microscopy (SEM): SEM is a powerful imaging technique used to visualize the surface morphology of materials at high magnification. It provides detailed information about surface topography, including features such as roughness, texture, and surface defects.
  • Energy-Dispersive X-ray Analysis (EDX): Coupled with SEM, energy-dispersive X-ray analysis (EDX) is utilized to determine the elemental composition of the material's surface. By detecting characteristic X-rays emitted from the sample when bombarded with an electron beam, EDX can identify the presence of different elements on the surface.
These surface characterization techniques are essential for assessing the surface properties, morphology, composition, and elemental distribution of materials, ensuring the quality, functionality, and performance of medical devices.